Date: 03/03/2019 Instrument configuration file: m:\PIRATA-FR29\data-processing\CTD\data\raw\fr29002.xmlcon Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Frequency channels suppressed : 0 Voltage words suppressed : 1 Computer interface : IEEE-488 (GPIB) Deck unit : SBE11plus Firmware Version >= 5.0 Scans to average : 1 NMEA position data added : Yes NMEA depth data added : No NMEA time added : Yes NMEA device connected to : PC Surface PAR voltage added : No Scan time added : Yes 1) Frequency 0, Temperature Serial number : 6083 Calibrated on : 06-Jun-18 G : 4.35319981e-003 H : 6.68879170e-004 I : 2.68418148e-005 J : 2.12681557e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 4509 Calibrated on : 07-Jun-18 G : -1.00364060e+001 H : 1.51500525e+000 I : -4.43743566e-004 J : 1.14695984e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 1263 Calibrated on : 15-Dec-15 C1 : -4.023025e+004 C2 : -4.466859e-001 C3 : 1.240000e-002 D1 : 3.511700e-002 D2 : 0.000000e+000 T1 : 3.023099e+001 T2 : -4.469710e-004 T3 : 4.128850e-006 T4 : 2.885040e-009 T5 : 0.000000e+000 Slope : 1.00000000 Offset : 0.00000 AD590M : 1.279148e-002 AD590B : -9.405686e+000 4) Frequency 3, Temperature, 2 Serial number : 6086 Calibrated on : 06-Jun-18 G : 4.35947792e-003 H : 6.70575828e-004 I : 2.73048438e-005 J : 2.16054153e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 4510 Calibrated on : 07-Jun-18 G : -9.74550860e+000 H : 1.55873621e+000 I : -1.41241334e-003 J : 1.86472266e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Oxygen, SBE 43 Serial number : 3261 Calibrated on : 13-Jun-18 Equation : Sea-Bird Soc : 4.66180e-001 Offset : -4.99200e-001 A : -4.51440e-003 B : 2.31760e-004 C : -3.75310e-006 E : 3.60000e-002 Tau20 : 1.01000e+000 D1 : 1.92634e-004 D2 : -4.64803e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 7) A/D voltage 1, Oxygen, SBE 43, 2 Serial number : 3265 Calibrated on : 13-Jun-18 Equation : Sea-Bird Soc : 3.53880e-001 Offset : -5.25700e-001 A : -4.58240e-003 B : 2.66640e-004 C : -4.02980e-006 E : 3.60000e-002 Tau20 : 1.66000e+000 D1 : 1.92634e-004 D2 : -4.64803e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 8) A/D voltage 2, Transmissometer, WET Labs C-Star Serial number : CTS1827DR Calibrated on : 27/07/2017 M : 19.2254 B : -0.0692 Path length : 0.250 9) A/D voltage 3, Fluorometer, WET Labs ECO-AFL/FL Serial number : FLRTD-4707 Calibrated on : 24/04/2017 Dark output : 0.0590 Scale factor : 6.00000000e+000 10) A/D voltage 4, Fluorometer, Chelsea Aqua 3 Serial number : 088-081 Calibrated on : 2016/01/29 VB : 0.148700 V1 : 2.094600 Vacetone : 0.280900 Scale factor : 1.000000 Slope : 1.000000 Offset : 0.000000 11) A/D voltage 5, Free Scan length : 42